ThinPrep® UroCyte™
(For use with Urine Cytology Processing or Slide-based Molecular Testing)
 |
1. Collection
Collect urine directly into the ThinPrep UroCyte Urine Collection Kit, OR process urine fresh. (Note:
Fresh urine can be mixed with a 2:1 urine to PreservCyt® ratio and stored for up to 48 hours before processing.)
|
 |
2. Concentrate by Centrifugation
(600g for 10 minutes)
|
 |
3. Pour Off Supernatant and Resuspend Cell Pellet
Resuspension can be done on a vortexor or may be achieved by syringing the pellet back and forth with a plastic pipette.
|
 |
4. Add CytoLyt® Solution Wash
- Add 30ml of CytoLyt Solution
- Centrifuge
- Pour Off Supernatant
- Resuspend Cell Pellet
|
 |
5. Evaluate Cell Pellet Appearance
If cell pellet is not free of blood, add 30ml of CytoLyt and repeat from step 2.
6. Add Specimen to PreservCyt Solution Vial
Allow to stand in PreservCyt Solution for 15 minutes.
|
 |
7. Run on ThinPrep Processor using Sequence 5 (UroCyte)
(Fix, stain, and evaluate cytology, or perform molecular diagnostic testing according to the manufacturer's instructions for use.)
|
Refer to the ThinPrep Processor Operator's Manual for detailed instrument and specimen preparation instructions.
ThinPrep UroCyte
Instructions for using the ThinPrep UroCyte Urine Collection Kit
Please refer to ThinPrep UroCyte PreservCyt Instructions for Use for complete instructions and information on limitations, warning, and precautions associated with the use of PreservCyt.
P/N 86369-002 Rev. C
|
|
 |
   |